Description
Raman
spectroscopy is the "fingerprint spectrum" of molecular vibration.
Different material molecules have different vibration frequencies, so
they are often used as an important basis for material identification.
Raman technology has many unique advantages over traditional infrared
and chemical methods. First, Raman scattering of water is very weak, so
Raman spectroscopy is an ideal tool for studying biological samples and
chemical compounds in aqueous solution. Second, Raman's peaks are sharp
and clear, and are more suitable for quantitative research, database
search, and qualitative analysis using differential analysis. Third,
Raman can cover about 4000 wavenumber intervals at the same time, and
can analyze organic matter and inorganic matter. If the infrared
technology covers the same Intervals must change gratings, beam
splitters, filters, and detectors. Fourth, samples (solid, liquid, gas)
for Raman measurements no need to be pretreated, and with advantages of
no contact, lossless, real-time and testing materials through
transparent packaging. The BRS-532 series Raman spectrometers are
equipped with a narrow linewidth low power laser with an excitation
wavelength of 532 nm. The spectral range is up to 4000 cm-1 and the
spectral resolution is up to 8 cm-1. According to the intensity of the
Raman spectrum, different sensitivity specifications can be selected,
which are BRS-532-01 universal type and BRS-532-02 high sensitive type.
The instrument is stable in performance, easy to carry, and supports OEM
customization and secondary development, providing a great convenience
for laboratory and on-site Raman detection.
Applications
Food safety
Jewelry appraisal
Chemicals, explosive materials and drug identification
Biomedical testing
Customs verification
Other applications where Raman is required.
Specifications
Model |
BRS-532-01 |
BRS-532-02 |
Dimension |
342*180*89mm |
342*180*89mm |
Weight |
2.45Kg |
2.45Kg |
Range |
200cm-1 - 4000cm-1 |
200cm-1 - 4000cm-1 |
Laser |
532nm±0.5nm
Line width<0.1nm
|
532nm±0.5nm
Line width<0.1nm
|
Resolution |
< 10cm-1 |
< 8cm-1 |
Fiber connector |
SMA905 |
SMA905 |
Laser Power |
0-500mW adjustable |
0-500mW adjustable |
Laser output port |
FC |
FC |
Probe focal length |
7.5mm |
7.5mm |
Detector |
Hamamatsu S11639
2048 Line array CMOS
|
Hamamatsu S11510
Area array back-thinned CCD
|
Signal to noise ratio |
600:1 full range |
800:1 full range |
Signal input port |
SMA905 |
SMA905 |
Integration time |
0.5ms - 10s |
0.5ms - 10s |
Power adapter |
100-240v AC 50/60Hz |
100-240v AC 50/60Hz |
Humidity |
5%-80% |
5%-80% |
Operating temperature
|
0℃ -45℃
(25℃ recommend)
|
0℃ -45℃
(25℃ recommend)
|
Signal output |
USB2.0,12Mbps |
USB2.0,12Mbps |